Level 1: Required Minimum information. It can be identified by cardinality no less than 1 and by having a green background.

Level 2: Recommended. It can be identified by cardinality starting from 0 and by having a yellow background.

Level 3: Optional. It can be identified by cardinality starting from 0 and by having a red background.

No.PropertyDefinitionExport FormatCardinalityHuman-Readable ExampleMachine-Readable ExampleComments
1Ultra-Violet-Visible Absorption Spectrophotometry Assayultraviolet-visible spectrophotometry (CHMO:0000292)1
1.1Ultra-Violet-Visible Absorption Spectrophotometry SampleA sample that has been investigated by a chemical analysis by UV-Vis absorption spectrophotometry1By a UV-Vis sample, we mean the last sample ready for placing in the instrument, i.e., it will already have the UV-Vis solvent added along with other possible additions.
1.1.1Characterized Compound
and/or
link to sample preparation resource
A compound characterized as part of a UV-Vis absorption Spectrophotometry sample using data acquired from a UV-Vis Absorption Sample analysis.
and/or
a link to the preparation steps of the UV-Vis Absorption sample.
  • Molfile (EDAM:3815)
  • ChEBI URL for a leaf frommolecular entity (CHEBI:23367)
  • PubChem URL
  • URL (EDAM:1052)
  • 1-ncaffeine CHEBI:27732If the compound is known, reporting it becomes mandatory. Repositories implementing this MIChI must provide a notice clarifying this to the users.
    1.1.1.1ConcentrationThe concentration of the Characterized Compound in the sample.A number with a unit that is a leaf from:
  • molar based unit UO:1000062
  • mass per unit volume based unit (UO:1000180)
  • volume per unit volume UO:0000205
  • 0-n0.5 molar0.5UO:0000062
    1.1.2UV-Vis SolventA solvent used to prepare the UV-Vis Absorption sample.ChEBI URL for a molecular entity that has therole solvent (CHEBI:46787)1-nWaterCHEBI:15377For solid UV-Vis, the solvent can be reported as none.
    1.1.2.1UV-Vis Solvent Volume/Volume Ratio Volume/volume ratios (SNOMED:282380000) of UV-Vis SolventA number without unit1-nWater: 90CHEBI:15377: 90
    1.1.3Cell path lengthcell path length (AFR:0001997)A number with a unit that is a leaf frommeter based unit (UO:1000008)110 mm10UO:0000016
    1.2Acquisition ParametersParameters used in the UV-Vis Absorption Spectrophotometry1
    1.2.1Sample Temperaturesample temperature (AFR:0002149)A number with a unit:kelvin (UO:0000012)1298.15 Kelvin298.15UO:0000012
    1.2.2Start of wavelength rangeThe shortest wavelength defined in the measurement scan, from which the spectrometer begins recording the absorbance of the sampleA number with a unit:
  • Wavelength (λ)nanometer (UO:0000018)
  • Frequencyhertz based unit UO:1000106
  • Wavenumber cm⁻¹reciprocal of (RO:0015012)centimeter UO:0000015
  • 0-1200 nm200UO:0000018
    1.2.3End of wavelength rangeThe longest wavelength defined in the measurement scan, at which the spectrometer stops recording the absorbance of the sample.A number with a unit:
  • Wavelength (λ)nanometer (UO:0000018)
  • Frequencyhertz based unit UO:1000106
  • Wavenumber cm⁻¹reciprocal of (RO:0015012)centimeter UO:0000015
  • 0-1400 nm400UO:0000018
    1.2.4Data IntervalThe instrument setting determining the wavelength increment between consecutive data points collected during a spectral scan.A number with a unit:
  • Wavelength (λ)nanometer (UO:0000018)
  • Frequencyhertz based unit UO:1000106
  • Wavenumber cm⁻¹reciprocal of (RO:0015012)centimeter UO:0000015
  • 0-11 nm1 UO:0000018Please note that this is the user input, not the device's maximum possible resolution. It has a synonym: resolution.
    1.2.5Bandwidthdetection bandwidth (AFQ:0000198)A number with a unit:
  • Wavelength (λ)nanometer (UO:0000018)
  • Frequencyhertz based unit UO:1000106
  • Wavenumber cm⁻¹reciprocal of (RO:0015012)centimeter UO:0000015
  • 0-11 nm1 UO:0000018
    1.2.6Scan RateThe speed at which a UV-Vis spectrophotometer scans across a range of wavelengths when collecting spectral data.A number with a unit: :rate unit (UO:0000280)
  • Nanometer per second (nm/s)
  • Frequencyhertz based unit UO:1000106
  • reciprocal centimeter per second cm^-1/s
  • 0-1240 nm/min240 nm/min
    1.2.7UV-Vis Average Time (or Averaging Time)The duration over which the instrument collects and averages the signal for each wavelength point during a spectral scan.A number with a unit that is a leaf from: time unit (UO:0000003)0-11 second1 UO:0000010
    1.2.8Beam ModeThe setup configuration of the instrument’s light path, impacting how the light interacts with the sample and reference.Free text. CV in preparation https://github.com/NFDI4Chem/workshops/issues/190-1Single-Beam ModeSingle-Beam Mode
    1.2.9Number of averagesnumber of measurement averages (AFR:0001673)A number with a unit: average0-13 averages3 averages
    1.3Instrumentultraviolet–visible spectrometer (CHMO:0001803)0-1
    1.3.1Instrument ManufacturerDevice Manufacture (NCIT:C156625)Free text. CV in preparation https://github.com/NFDI4Chem/workshops/issues/190-1Agilent TechnologiesAgilent Technologies
    1.3.2Instrument ModelInstrument Model (NCIT:C177610)Free text. CV in preparation https://github.com/NFDI4Chem/workshops/issues/190-1Agilent Cary 60Agilent Cary 60
    1.3.3Light sourcelight source (OBI:0400065)Free text. CV in preparation https://github.com/NFDI4Chem/workshops/issues/190-nDeuterium lampDeuterium lamp
    1.4Acquisition SoftwareSoftware related to the recording of spectra.0-1
    1.4.1Software namesoftware name (AFR:0002802)Free text. CV in preparation https://github.com/NFDI4Chem/workshops/issues/190-1Cary WinUVCary WinUV
    1.4.2Software versionsoftware version (AFR:0001700)Free text. CV in preparation https://github.com/NFDI4Chem/workshops/issues/190-15.35.3
    1.5Data Processing1
    1.5.1X axis LabelThe label of the X axis in the UV-Vis spectrum
  • Wavelength (λ)wavelength (quality) PATO:0001242
  • Frequencyfrequency AFR:0001732
  • Wavenumberwavenumber AFR:0001592
  • 1Wavelength (λ)Wavelength (λ)
    1.5.2X axis UnitThe unit of the X axis in the UV-Vis spectrum
  • Wavelength (λ):nanometer (UO:0000018)
  • Frequency:frequency AFR:0001732
  • Wavenumber: cm⁻¹
  • 1nmUO:0000018
    1.5.3Y axis LabelThe label of the Y axis in the UV-Vis spectrum
  • Absorbance (A)AFQ:0000061Absorbance Units (AU)
  • Molar absorption coefficient (ε)
  • Log Absorbance (Log A)
  • Intensity
  • 1Molar absorption coefficient (ε)Molar absorption coefficient (ε)
    1.5.4Y axis UnitThe unit of the Y axis in the UV-Vis spectrum
  • Absorbance (A) / Absorbance Units (AU): null (dimensionless)
  • Molar absorption coefficient (ε): L⋅mol−1⋅cm−1
  • Log Absorbance (Log A): null
  • Intensity: W/cm2
  • 1L⋅mol−1⋅cm−1L⋅mol−1⋅cm−1
    1.5.5Baseline Correction AlgorithmThe name of the algorithm used for Baseline Correction.Free text. CV in preparation https://github.com/NFDI4Chem/workshops/issues/190-1Linear Baseline CorrectionLinear Baseline Correction
    1.5.5.1Baseline Correction ParameterA parameter to a baseline correction function.Free text0-n
    1.5.6Smoothing AlgorithmThe name of the algorithm used for SmoothingFree text. CV in preparation https://github.com/NFDI4Chem/workshops/issues/190-1Moving AverageMoving Average
    1.5.6.1Smoothing parametersA parameter to a smoothing correction function.Free text0-n